CONDITIONS
column
Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)
gradient
5 mM ammonium formate in acetonitrile:water (90:10, v/v)
flow rate
0.4 mL/min
column temp.
35 °C
detector
TOF/MS
injection
2 μL
sample
300 ng/mL in acetonitrile
Beschreibung
Hinweis zur Analyse
The fluoroamphetamines (FA) are a group of "designer" stimulant drugs in amphetamine family. The rapid LC/MS analysis of four members of the FA family on Ascentis Express HILIC is shown here. Cerilliant CRMs were used to ensure reliable MS identification and quantification.