HPLC Analysis of Diet Soda Spiked with Neotame on Ascentis® Express C18
Materialien
Analytische Säule
Produkt-Nr.
Beschreibung
Preisangaben
CONDITIONS
column
Ascentis Express C18, 3 cm x 4.6 mm I.D., 2.7 μm particles (53818-U)
mobile phase
[A] 100 mM ammonium acetate, pH 5.6 (titrated with acetic acid); [B] water; [C] acetonitrile
gradient
20% A constant; 75 to 20% B, 5 to 60% C in 1 min; held at 20% B, 60% C for 0.1 min
flow rate
3 mL/min
pressure
3191 psi (220 bar)
column temp.
40 °C
detector
UV, 214 nm
injection
1 μL
sample
neotame 350 μg/mL in diet cola
Beschreibung
Hinweis zur Analyse
This application demonstrates the suitability of Ascentis Express C18 for the efficient separation of common drink additives used in diet soda.
Rechtliche Hinweise
Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany