LC/MS/MS Analysis of Pesticide Residues in Pistachios on the Ascentis® Express RP-Amide Column after QuEChERS Extraction

LC/MS/MS Analysis of Pesticide Residues in Pistachios on the Ascentis® Express RP-Amide Column after QuEChERS Extraction application for LC-MS, application for SPE

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Ascentis® Express RP-Amide, 2 μm UHPLC Column

2 μm particle size, L × I.D. 10 cm × 2.1 mm

Préparation d'échantillons

Réf. du produit
Description
Tarif

Supel QuE

Citrate Extraction Tube, pk of 50, suitable for EN 15662

Supel QuE

Empty Centrifuge Tube with Lid, centrifuge tube volume 50 mL, pk of 50, suitable for EN 15662:2008 per BS

Benchmixer XL multi-tube vortexer

AC/DC input 115 V AC

Supel QuE

Z-Sep+ Tube, centrifuge tube volume 2 mL, pack of 100 ea

QuEChERS Shaker and Rack Starter Kit

AC/DC input 115 V AC (USA Compatible Plug)

Tube ou plaque pour SPE

Réf. du produit
Description
Tarif

Supel QuE

Z-Sep+ Tube, pk of 50, centrifuge tube volume 15 mL , Shaker Compatible

À utiliser avec

Réf. du produit
Description
Tarif

Certified Vial Kit, Low Adsorption (LA), 2 mL, pk of 100

volume 2 mL, amber glass vial (with marking spot), natural PTFE/silicone septa (with slit), thread for 9 mm

CONDITIONS

sample/matrix

5 g ground pistachios spiked with pesticides at 10 ng/g

extraction process

add 10 mL water and 10 mL acetonitrile; shake for 1 hour at 2500 rpm; add contents of Supel QuE citrate extraction tube; shake for 1 minute; centrifuge at 5000 rpm fo 5 minutes; draw off supernatant.

clean-up tube

Supel QuE Z-Sep + (55408-U)

clean-up process

transfer 1 mL of extract to 2 mL tube containing 75 mg Z-Sep+; shake for 1 minute; centrifuge at 5000 rpm for 3 minutes; draw off supernatant.

sample preparation

Dilute 100 μL of cleaned extract to a final volume of 1 mL in 5 mM ammonium formate/0.1% formic acid in water. Transfer diluted sample to vial for LC/MS/MS analysis

column

Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2 μm particles (51576-U)

mobile phase

[A] 5 mM ammonium formate + 0.1% formic acid in water; [B] 5 mM ammonium formate + 0.1% formic acid in 95:5 acetonitrile:H2O

gradient

95% A from 0 to 1 min; to 100% B in 13 min; held at 100% B for 1.5 min; to 95% A in 0.5 min; held at 95% A for 1.5 min

flow rate

0.4 mL/min

column temp.

30 °C

detector

MS, ESI (+), MRM

injection

5 μL

extraction tube

PK50 Supel QuE Citrate (EN) Tube (55227-U)

Description

Remarque sur l'analyse

The quick, easy, cheap, effective, rugged, and safe (QuEChERS) method can be used to analyze multiple pesticide residues found in pistachios. Because they contain a significant amount of fat, a cleanup step should be used as part of the QuEChERS method. In this study, QuEChERS extraction and cleanup using Z-Sep+ was used for the analysis of pesticide residues of interest in pistachios. Average recoveries obtained for replicate samples spiked at 10 ng/g were in the range of 70-120% for 23 of the 25 pesticides analyzed. .

Autres remarques

Freeze whole pistachios (with shells) using liquid nitrogen, and grind.

Informations légales

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany