LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 4

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 4 application for HPLC

Matériaux

Application

Réf. du produit
Description
Tarif

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 1

application for HPLC

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 2

application for HPLC

LC/MS Analysis of Illicit Bath Salts on Ascentis® Express HILIC, Set 3

application for HPLC

Colonne analytique

Réf. du produit
Description
Tarif

Ascentis® Express HILIC HPLC Column

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

Étalon

Réf. du produit
Description
Tarif

Buphedrone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Mephedrone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Methylone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

MDAI hydrochloride solution

100 μg/mL in acetonitrile: water (9:1) with 5% 1 M HCl (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Pyrovalerone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

N-Ethylcathinone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

5 mM ammonium formate in [A] acetonitrile; [B] water; (98:2, A:B)

flow rate

0.6 mL/min

pressure

267 psi (18.4 bar)

column temp.

35 °C

detector

MS, ESI(+) TOF, Summed Ion Chromatogram

injection

2 μL

sample

500 ng/mL in acetonitrile

Description

Remarque sur l'analyse

HILIC mode operation on an Ascentis Express column provided baseline resolution of these bath salts, including isobaric compounds. Ultra high purity solvents were used to supply low background interference and low particulate contaminants for robust; trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.