HPLC Analysis of Withania Extract on Ascentis® Express Phenyl-Hexyl
Matériaux
Colonne analytique
Réf. du produit
Description
Tarif
Ascentis® Express Phenyl-Hexyl, 2.7 μm HPLC Column
2.7 μm particle size, L × I.D. 10 cm × 2.1 mmCONDITIONS
column
Ascentis Express Phenyl-Hexyl, 10 cm x 2.1 mm I.D., 2.7 μm particles (53336-U)
mobile phase
[A] water; [B] acetonitrile
gradient
20 to 100% B in 10 min; held at 100% B for 0.5 min
flow rate
0.3 mL/min
pressure
160 bar (2320 psi)
column temp.
35 °C
detector
UV, 227 nm
injection
5 μL
sample
20 mg/mL in water:acetonitrile (50:50); sonicate 15 minutes; filter 0.45 μm;
dilute to water:acetonitrile (80:20)
Description
Remarque sur l'analyse
This application demonstrates the suitability of Ascentis Express Phenyl-Hexyl for the efficient separation of withania extract.