Accéder au contenu
MilliporeSigma

AOAC Method 2012.13: GC Analysis of Fatty Acids in Powdered Infant Formula on SP®-2560

AOAC Method 2012.13: GC Analysis of Fatty Acids in Powdered Infant Formula on SP®-2560 suitable for GC

Matériaux

Colonne analytique

Réf. du produit
Description
Tarif

Colonne capillaire pour la GC SP®-2560

L × I.D. 100 m × 0.25 mm, df 0.20 μm

Étalon

Réf. du produit
Description
Tarif

Methyl undecanoate

analytical standard

CONDITIONS

sample/matrix

weigh 168 mg powdered infant formula (approximately 50 mg of fat) into a 25 mL centrifuge tube; add 2 mL of water and mix to dissolve; (let sit for 15 minutes at room temperature; add 5 mL of internal standard (C11:0 FAME + C13:0 TAG, each at 2 mg/mL in methyl tert butyl ether))

extraction

5 mL of 5% (w/v) methanolic sodium methoxide solution; close tube and vortex for 10 seconds; after 180 seconds (time starts when sodium methoxide is added) add 2 mL of hexane; (after 210 seconds add 10 ml of neutralization solution (10% disodium hydrogen citrate/15% sodium chloride in water); gently shake using vortex mixer; centrifuge mixture at 1750 rpm for 5 minutes)

eluate post-treatment

transfer 200 μL of supernatant into 10 mL flask and dilute to mark with hexane

column

SP-2560, 100 m x 0.25 mm I.D., 0.20 μm (24056)

oven

60 °C (1 min), 15 °C/min to 165 °C (1 min), 2 °C/min to 225 °C (20 min)

inj. temp.

250 °C

detector

FID, 250 °C

carrier gas

helium, 0.8 mL/min

injection

1μL, 10:1 split

liner

4 mm I.D., split/splitless type, wool packed single taper FocusLiner design

Description

Remarque sur l'analyse

The nutritional content of infant formula is extremely critical, and thus is one of the most regulated food products worldwide. Most commercial infant formulas are derived from cow′s milk, and are formulated to mimic human breast milk. They offer the advantage of being nutritionally balanced and easy for most babies to digest. All infant formulas, whether cow′s milk-based or from another source, are subject to specific regulatory nutritional content requirements. AOAC Method 2012.13 using a 100 m SP-2560 is able to provide identification of fatty acids, as well as determination of total fatty acid content and breakdown by class.

Informations légales

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
SP is a registered trademark of Sigma-Aldrich Co. LLC