Determination of Ephedrine HCl and Pseudoephedrine HCl in Xiao’er Kechuanling Oral Solution with Purospher® STAR RP-18e colμmn
Dean Duan
MilliporeSigma China Application Lab Shanghai, China
Chinese Pharmacopeia Method
Ephedrine hydrochloride and pseudoephedrine hydrochloride can be used to reduce cold, allergic rhinitis, rhinitis and sinusitis caused by nasal congestion symptoms and control bronchial asthma. The same compounds are also found in the Chinese traditional medicine Xiao’er Kechuanling oral solution.
This application focuses on the determination of ephedrine HCl and pseudoephedrine HCl. Oral solution samples were distilled to extract the ephedrine HCl and pseudoephedrine HCl. Samples were then filtered through Millex® PTFE syringe filters, prior to HPLC-UV analysis using a Purospher® STAR RP-18 endcapped HPLC colμmn.
The limit of detection (LOD) and the limit of quantitation (LOQ) were 0.04 µg/kg and 0.12 µg/kg, respectively, for ephedrine, and 0.02 µg/kg and 0.06 µg/kg, respectively, for pseudoephedrine. The method can be used for determination of ephedrine HCl and pseudoephedrine HCl in Xiao’er Kechuanling oral solution.
![Chemical structures of ephedrine HCl and pseudoephedrine HCl 2-D chemical structures (bond line structure) of ephedrine HCl and pseudoephedrine HCl compounds used to determine analytes in Xiao’er Kechuanling oral solution](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/chemical-structures-of-ephedrine-hci-and-pseudoephedrine-hci/chemical-structures-of-ephedrine-hci-and-pseudoephedrine-hci.jpg)
Figure 1.Chemical structures of compounds used in the study.
Determination of Ephedrine HCl and Pseudoephedrine HCl in Xiao’er Kechuanling Oral Solution
Purospher® STAR RP-18e
![HPLC profile of ephedrine HCl and pseudoephedrine HCl standards Separation of ephedrine HCl and pseudoephedrine HCl standards used to determine analytes in Xiao’er Kechuanling oral solution by HPLC-UV method](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/hplc-profile-of-ephedrine-hci-and-pseudoephedrine-hci-standards/hplc-profile-of-ephedrine-hci-and-pseudoephedrine-hci-standards.jpg)
Figure 2.Chromatogram of ephedrine HCl and pseudoephedrine HCl standard solution.
![HPLC profile of Xiao’er Kechuanling oral solution Chromatogram showing standard peaks of Xiao’er Kechuanling oral solution used to determine analytes in Xiao’er Kechuanling oral solution by HPLC-UV method](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/hplc-profile-of-xiao-er-kechuanling-oral-solution/hplc-profile-of-xiao-er-kechuanling-oral-solution.jpg)
Figure 3.Chromatogram of in Xiao’er Kechuanling oral solution.
![HPLC of solvent blank Chromatographic run of solvent blank in HPLC-UV separation of ephedrine HCl and pseudoephedrine HCl standards used to determine analytes in Xiao’er Kechuanling oral solution](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/hplc-of-solvent-blank-hci/hplc-of-solvent-blank-hci.jpg)
Figure 4.Chromatogram of solvent blank.
Specificity and Repeatability – Ephedrine and Pseudoephedrine
Specificity: Inject reference solution and determine the retention time of desired analyte in presence of other components like impurities. |
---|
Calibration Data - Ephedrine
Linearity (area mAU*min), ephedrine HCl |
---|
LOD and LOQ, ephedrine HCl |
---|
![Calibration graph of ephedrine HCl Calibration curve of ephedrine HCl used to measure the analyte in Xiao’er Kechuanling oral solution using Purospher® STAR RP-18e colμmn](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/calibration-graph-of-ephedrine-hci/calibration-graph-of-ephedrine-hci.jpg)
Figure 5.Calibration curve of ephedrine HCl.
Calibration Data - Pseudoephedrin
Linearity (area mAU*min), pseudoephedrine HCl |
---|
4.2 LOD and LOQ - pseudoephedrine HCl |
---|
![Calibration graph of pseudoephedrine HCl Calibration curve of pseudoephedrine HCl to measure the analyte in Xiao’er Kechuanling oral solution using Purospher® STAR RP-18e colμmn](/deepweb/assets/sigmaaldrich/marketing/global/images/technical-documents/articles/analytical-chemistry/small-molecule-hplc/calibration-graph-of-pseudoephedrine-hci/calibration-graph-of-pseudoephedrine-hci.jpg)
Figure 6.Calibration curve of pseudoephedrine HCl.
Conclusion
The method can be used for ephedrine HCl and pseudoephedrine HCl in Xiao’er Kechuanling oral solution.