parâmetros de teste
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: Fresh 1 L water samples, dechlorinated with sodium thiosulfate when necessary, at ambient temperature and pH.
SPE tube/cartridge: Supelclean ENVI-Carb, 250 mg/6 mL (57092)
condition: 10 mL DI water at 20 mL/min
sample addition: 0.9 L water sample at 20 mL/min
drying: 10 min using clean nitrogen
washing: 10 mL DI water at 20 mL/min
elution: 10 mL 0.1% phosphoric acid in methylene chloride:acetonitrile (80:20) at 5 mL/min
column: polymeric-coated silica-based PAH specialty column, 20 cm × 3 mm I.D., 5 μm (Supelco equivalent, SUPELCOSIL LC-PAH, available upon request)
mobile phase: gradient, [A]: 0.05% phosphoric acid in DI water; [B]: acetonitrile
flow rate: 0.5 mL/min
column temp.: 50 °C
detector: photodiode array- peak width: 0.053 min, sampling interval: 0.320 sec, monitor 210 nm & 225 nm
injection: 10 μL of extract (4-5 ppb each analyte in water)
adequação
application for HPLC
application for SPE
aplicação(ões)
environmental
Aplicação
Informações legais
Analito
coluna analítica
Escolha uma das versões mais recentes:
Certificados de análise (COA)
Lamentamos, não temos COA para este produto disponíveis online no momento.
Se precisar de ajuda, entre em contato Atendimento ao cliente
Já possui este produto?
Encontre a documentação dos produtos que você adquiriu recentemente na biblioteca de documentos.
Nossa equipe de cientistas tem experiência em todas as áreas de pesquisa, incluindo Life Sciences, ciência de materiais, síntese química, cromatografia, química analítica e muitas outras.
Entre em contato com a assistência técnica