Analytical Method for Selenium in Water
This method for measuring selenium in water uses the Spectroquant® Photometer and a clear, filtered sample to evaluate the results at 445 nm.
Instruments and Reagents Required
Spectroquant® Photometer NOVA 30 (Product No. 109748) or
Spectroquant® Photometer NOVA 60 (Product No. 109751)
Rectangular cells 50 mm (Product No. 114944)
Formic acid 98-100% for analysis (Product No. 100264 )
3,3‘-Diaminobenzidine tetrahydrochloride (CAS 868272-85-9) (Product No. 32750)
TitriplexIII® solution 0.1 mol/l (Product No. 109992)
Ammonia solution 25% for analysis (Product No. 105432)
Toluene for gas chromatography (Product No. 108389)
pH-indicator strips, pH 2.0 - 9.0 (Product No. 1.09584)
Sodium sulfate anhydrous for analysis (Product No. 106649)
Water for analysis (Product No. 116754)
Membrane filter with 0.45 μm pore diameter
Measuring Range
0.05 – 2.00 mg/L
Reagent 1
Fill 9.4 mL formic acid 100% in a calibrated or conformity-checked 100-mL volumetric flask. Make up to volume with water for analysis and mix.
Reagent 2
Dissolve 0.5 g 3,3’-diaminobenzidine tetrahydrochloride in 100 mL chilled water for analysis. Prepare freshly each day and store in an amber glass bottle.
Reagent 3
Fill 48 mL ammonia solution 25% for analysis in a calibrated or conformity-checked 100-mL volumetric flask. Make up to volume with water for analysis and mix.
Sample Preparation
The sample must be clear; if not, filter through a membrane filter with 0.45 μm pore diameter.
Analysis
- Mix 30 mL of nearly neutral sample solution with 2 mL reagent 1, 2 mL Titriplex III® solution 0.1 mol/l and 2 mL reagent 2.
- Adjust the pH value of the mixture with reagent 3 to 2.5 and set aside for 45 min in the dark.
- Adjust the pH value of the mixture again with reagent 3 to 7.0, transfer into a 100-mL separation funnel, add 15 mL water for analysis and 7 mL toluene for analysis. Shake the mixture in the closed funnel for 1 min.
- Allow to separate and discharge the lower water phase.
- Fill the toluene extract in a test tube, add some anhydrous sodium sulfate and shake.
- Fill the clear, dried toluene extract in a 50 mm rectangular cell and measure at 445 nm against pure toluene.
Parameter list: |
---|
Calculation when measuring the absorbance value using NOVA photometers without programming the method in the memory of the instrument:
Concentration [mg/l] = Absorbance x 0.654 – 0.037
Calculation:
Result = measured value in mg/l Se4+
Statistical Data:
Standard deviation: 0.019 mg/l