Skip to Content
Merck

GC Analysis of Pesticides in Spinach on Equity-1 after SPE using Supelclean ENVI-Carb-II/PSA

GC Analysis of Pesticides in Spinach on Equity-1 after SPE using Supelclean™ ENVI-Carb-II/PSA suitable for GC, application for SPE

Materials

analytical column

Product No.
Description
Pricing

Equity-1 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.25 μm

SPE tube or plate

Product No.
Description
Pricing

Supelclean ENVI-Carb II/PSA SPE Tube

bed A 500 mg (Supelclean Envi-Carb-II SPE), bed B 300 mg (Supelclean PSA SPE), volume 6 mL, pk of 30

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

10 g of fresh spinach spiked at 0.2 ppm each pesticide

SPE tube/cartridge

Supelclean ENVI-Carb-II/PSA, 500mg/300mg/6mL (55119-U)

condition

5 mL acetonitrile:toluene (3:1, v/v)

sample addition

1 mL concentrated sample extract

elution

20 mL acetonitrile:toluene (3:1, v/v)

eluate post-treatment

Evaporate eluate to dryness with nitrogen at 40 °C. Reconstitute in 1 mL acetone:hexane (1:1, v/v).

column

Equity-1, 30 m x 0.25 mm I.D., 0.25 μm (28046-U)

oven

50 °C (5 min), 25 °C/min to 125 °C, 10 °C/min to 30 °C (8 min)

inj. temp.

200 °C, aux: 325 °C (8 min)

detector

MSD, scan range 45-450 amu

carrier gas

helium, 0.9 mL/min., constant flow mode

injection

1 μL, splitless (splitter open at 1 min)

liner

4 mm I.D., single taper

Description

Application

Sample Pre-treatment:
Homogenize 10 g of spinach in 10 mL of acetonitrile.

Legal Information

null
Supelclean is a trademark of Sigma-Aldrich Co. LLC